Technology: ST 130nm BiCMOS9MW
Frederik Dreyer, Daniel Krüger, Jens Anders
Increasing sensitivity and signal robustness
The presented chip was designed to conduct in-vivo magnetic resonance (MR) experiments inside rat brains and to offer superior sensitivity and signal robustness compared to extracorporeal detection coils and passive implantable MR coils.
Designing the NMR-Needle
The so-called NMR-on-a-chip approach [1] suggests co-integrating a miniaturized MR detection coil with the complete MR transceiver on a single implantable sensor.
The on-chip MR coil is implemented in the two thick top metal layers provided by ST’s BiCMOS9MW technology. By connecting multi-turn coils in the two layers in series, the induced MR signal can be maximized to alleviate the burden on the receiver electronics.
To allow for implantation with minimum tissue damage, the dies are post-processed by mechanically thinning them down to 80µm and producing a tip at an optimized angle by wafer dicing.
Improving performance and application range
This chip is the next generation of the device presented in [2], further improving its performance and application range in the target MR application as well as minimizing tissue damage during the implantation surgery by using an optimized chip geometry. Applications of the device include, for instance, the detection of changes in blood flow and oxygenation as well as in-vivo MR imaging and heteronuclear MR spectroscopy.